1. When the autosampler and thermostatic autosampler are powered on, never disconnect or reconnect the cable between these two components. This will destroy the circuit of the module.
2. Unplug the power cord from the autosampler and thermostat to disconnect the autosampler from the line power supply. However, even if the power switch on the front panel of the autosampler is turned off, the autosampler is still charged. Please make sure that you can unplug the power plug at any time.
3. If the equipment is connected to a line voltage exceeding the specified line voltage, it may cause electric shock or damage to the instrument.
4. Make sure that the condensate pipe is always above the liquid level of the container. If the condensate pipe extends into the liquid, the condensate cannot flow out of the pipe and blocks the outlet. This will damage the circuit of the instrument.